Home Categories Send inquiry

New Generation Intelligent Electrosurgical Generator

Features

Two monopolar output ports 4 monopolar cutting modes: pure cut, blend 1, blend 2, blend 3 Pure cut: cut the tissue cleanly and accurately without coagulation blend 1: Use when the cutting speed is slightly slow and a small amount of hemostasis is required. blend 2: Compared with blend 1, it is used when the cutting speed is slightly slower and a better hemostatic effect is needed. blend 3: Compared with blend 2, it is used when the cutting speed is slower, and a much better hemostatic effect is needed. 3 coagulation modes: spray coagulation, forced coagulation, and soft coagulation spray coagulation: high-efficiency coagulation without contact surface. The coagulation depth is shallow. The tissue is removed by evaporation. It usually uses a Blade or ball electrode for coagulation. forced coagulation: It’s non-contact coagulation. The output threshold voltage is lower than spray coagulation. It’s suitable for coagulation in a small area. soft coagulation: Mild coagulation penetrates deeply to prevent tissue carbonization and reduce electrode adhesion to tissue. 3 bipolar output modes: macro mode, standard mode, and fine mode Macro mode: It’s used in bipolar cutting or rapid coagulation. Voltage is higher and the power is higher than standard and fine mode. Standard mode: It’s suitable for most bipolar applications. Keep the low voltage to prevent sparks. Fine mode: It’s used for high precision and fine control of drying amount. Keep the low voltage to prevent sparks. CQM contact quality monitoring system Automatically monitor the quality of contact between the dispersive pad and the patient in real-time. If the contact quality is lower than the set value, there will be a sound and light alarm and cut off the power output to ensure safety. allow two electrosurgical pencils to cut and coagulate simultaneously 2 control way-Electrosurgical pens and foot switch control Start with the recently used mode, power, and other parameters 9 sets of memory modes, power parameters, etc., can be quickly recalled. Volume adjustment function Cut and coagulate in an intermittent manner

Key Specifications

Mode

Max Output Power(W)

Load impedance (Ω)

Modulation Frequency (kHz)

Max Output Voltage (V)

Crest Factor

Monopolar

Cut

Pure Cut

300

500

——

1050

1.3

Blend 1

250

500

25

1350

1.6

Blend 2

200

500

25

1200

1.6

Blend 3

150

500

25

1050

1.6

Coag

Spray

120

500

25

1400

2.4

Forced

120

500

25

1400

2.4

Soft

120

500

25

1400

2.4

Bipolar

Marco

150

100

——

450

1.5

Standard

100

100

——

400

1.5

Fine

50

100

——

300

1.5


New Generation Intelligent Electrosurgical Generator Related Video: